Acta Mechanica Slovaca 2026, 30(1):58-63 | DOI: 10.21496/ams.2026.013
Metrological Comparison of Ocular Tonometers for Optometric Use
- 1 Department of Optics, Faculty of Science, Palacký University, 17. listopadu 12, 771 46 Olomouc, Czech Republic
- 2 Unit of Primary Metrology of Pressure, Vacuum and Low Mass Flow, Czech Metrology Institute, Okružní 31, 638 00 Brno, Czech Republic
This paper presents an international metrological comparison of non-contact tonometers used for measuring intraocular pressure (IOP), a critical parameter in glaucoma diagnosis. Since non-contact tonometry is affected by biological variability and lacks direct linkage to SI units, reliable calibration standards are essential. A newly developed universal model eye, independent of specific manufacturers and traceable to SI units, served as the primary metrological standard. The comparison involved tonometers tested at institution in the Czech Republic and Slovakia according to the ISO 8612:2009 standard under controlled laboratory conditions. Accuracy, repeatability, and deviations from reference values were evaluated. Results demonstrated stable performance with minimal deviations across the measured pressure range. The study confirms the suitability of the developed model eye as a universal calibration standard and highlights the importance of international comparisons for enhancing the reliability and standardization of IOP measurements in clinical practice.
Keywords: Intraocular pressure, Metrological comparison, Metrological traceability, Ocular tonometers
Received: February 15, 2026; Revised: March 11, 2026; Accepted: March 18, 2026; Published: April 1, 2026 Show citation
References
- Rybář, J.; Suchý, V.; Ďuriš, S.; Pavlásek, P.; Grosinger, P.; Hučko, B.; Ferková, S. L.; Sekáč, J.; Chytil, M. Metodika kalibrací bezkontaktních očních tonometrů. In Metrológia a skúšobníctvo. Roč. 25, č. 1 (2020), s. 8-13. /Methodology for the calibration of non-contact ocular tonometers. In Metrology and Testing. Vol. 25, No. 1 (2020), pp. 8-13/. ISSN 1335-2768.
- Rybář, J.; Leja, J.; Dunaj, Š.; Smetánka, A. and Onderčo, P. New design solution of reference device with model eye. In 10th International scientific conference on advances in mechanical engineering: Selected peer-reviewed full text papers from the 10th International scientific conference on advances in mechanical engineering (ISCAME 2024). 1. vydanie /1st edition/. Zurich, Švajčiarsko: Trans Tech Publications, 2025, S. /PP./ 81 - 86. ISSN 1662-0356. In the database: SCOPUS: 2-s2.0-105008444446.
- ISO 8612:2009. Ophthalmic Instruments - Tonometers. Prague: Office for Technical Standardization, Metrology and State Testing, 2009. 28 pages.
- Rybář, J.; Ďuriš, S.; Pavlásek, P.; Šrenkel, P. and Chytil, M. Měření nitroočního tlaku v oftalmologické praxi. In Metrológia a skúšobníctvo. Roč. 23, č. 1 (2018), s. 8-12. /Measurement of Intraocular Pressure in Ophthalmic Practice. In Metrology and Testing. Vol. 23, No. 1 (2018), pp. 8-12/. ISSN 1335-2768
- Morrison, J. C. and Pollack, I. P. Glaucoma: science and practice. [s.l.]: Thieme, 2003. 530 s. /530 pgs./.
Go to original source... - Stein JD, Khawaja AP, Weizer JS. Glaucoma in adults-screening, diagnosis, and management: a review. JAMA 2021;325:164-74. doi: 10.1001/jama.2020.21899
Go to original source... - Weinreb RN, Aung T, Medeiros FA. The pathophysiology and treatment of glaucoma: a review. JAMA 2014;311(18):1901-11. doi:10.1001/jama.2014.3192.
Go to original source... - Smetánka, A.; Rybář, J.; Onderčo, P.; Leja, J. and Gerneschová, J. What is metrological traceability and what role does physics play? In 27th Conference of Slovak Physicists: Proceedings. 1. vydanie /1st edition/. Košice: Slovak PhysicalSociety, 2024, S. 65 - 66. ISBN 978-80-89855-26-1. In the database: SCOPUS: 2-s2.0-85214019262.
- Kelemenová, T. and Dovica, M. Kalibrácia meradiel. 1. vydanie. Košice: Technická univerzita v Košiciach, Strojnícka fakulta. Edícia vedeckej a odbornej literatúry, 2016. 232 s. Calibration of Measuring Instruments. 1st edition. Košice: Technical University of Košice, Faculty of Mechanical Engineering. Scientific and Professional Literature Series, 2016. 232 pp./. ISBN 978-80-553-3069-3.
- RYBÁŘ, Jan. Development of methods for traceability assurance of intraocular pressure measuring instruments. [Dissertation thesis]. Slovak University of Technology in Bratislava, Faculty of Mechanical Engineering. Thesis supervisor: prof. Ing. Stanislav Ďuriš, PhD. Bratislava: SjF STU BA, 2019. 117p
- Rybář, J.; Hučko, B.; Ďuriš, S.; Pavlásek, P.; Chytil, M.; Furdová, A. and Veselý, P. Factors affecting measurements of IOP using non-contact eye tonometer. In Strojnícky časopis = Journal of Mechanical engineering. Roč. 70, č. 2 (2020), s. 133-140 /Vol. 70, No. 2 (2020), pp. 133-140/. ISSN 0039-2472 (2020). In the database: SCOPUS: 2-s2.0-85099079234.
Go to original source... - Tham, YC.; Li, X.; Wong TY.; et al. Global prevalence of glaucoma and projections of glaucoma burden through 2040: a systematic review and meta-analysis. Ophthalmology. 2014; 121: 2081-2090. DOI: 10.1016/j.ophtha.2014.05.013.
Go to original source...
This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), which permits use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.